450-500 Watt Single Output Medical Power Supplies : FPM500
FPM500 - 450-500 Watt Single Output Medical Power Supplies

Description
Medical Open Frame
The FPM500 series of AC-DC medical approved switching power supplies in a package of 4 x 7 x 1.7 inches are capable of delivering 450-500 watts of continuous power at 30 CFM forced air cooling or 350-400 watts at convection cooling. The units are constructed on a printed circuit board with a U-bracket for mechanical support and heat sinking. A cover and fan assembly can be added during manufacturing. They are designed for medical applications including those needing BF rated insulation and/or an operation altitude up to 5000 meters.Features
- BF Class insulation
- Operation altitude up to 5000 meters
- 100-240 VAC input with active PFC
- Less than 300 μA leakage current
- EN55011 /55022 Class B conducted emissions
- Inhibit - TTL high to disable output
- Compliant with RoHS requirements
- Power consumption in standby mode less than 1W at standby power 5 V /100 mA
Specification
Input Voltage | 90-264 VAC |
Input frequency | 47-63 Hz |
Input Current (rms) | 5.2 A (rms) @115 VAC, 60 Hz , 2.6 A (rms) @ 230 VAC, 50 Hz |
Earth leakage current | 300 μA max. @ 264 VAC, 63 Hz |
Touch current | 100 μA max. @ 264 VAC, 63 Hz |
Output voltage/current | See rating chart. |
Maximum output power | See rating chart. |
Ripple and noise | 1% peak to peak maximum |
Remote sense | Compensation for cable losses up to 0.5V |
Over power protection | Set at 105-140% of its maximum output power, Automatic recovery |
Over voltage protection | Set at 112-140% of its rated output voltage, latching by recycle input to reset |
Short circuit protection | Automatic recovery |
Over temperature protection | Latching by recycle input to reset |
Temperature coefficient | All outputs ±0.04% /℃ maximum |
Transient response | Maximum excursion of 4%, recovering to 1% of final value within 500 us after a 25% step load change |
Standby power | 5 V at 500 mA maximum |
Fan power | 12 V at 300 mA maximum |
Operating temperature | -10℃ to +70℃ |
Storage temperature | -40℃ to +85℃ |
Relative humidity | 5% to 95% non-condensing |
Temperature derating | Derate from 100% at +50℃ linearly to 50% at +70℃, applicable to convection and forced-air cooling conditions |
Switching frequency | 55-300 KHz |
Efficiency | Typical 90% |
Hold-up time | 20 ms minimum at 110 VAC & 500 W |
Line regulation | ±0.5% maximum at full load |
Inrush current | 30 A @ 115 VAC, or 60 A @ 230 VAC, at 25℃ cold start |
Withstand voltage | 4000 VAC from input to output (2 MOPP) , 1500 VAC from input to ground (1 MOPP) , 1500 VAC from output to ground |
MTBF | 100,000 hours at full load at 25℃ ambient, calculated per MIL-HDBK-217F EMC Performance |
EN55011/ EN55032 | Class B conducted, Class B radiated |
EN61000-3-2 | Harmonic distortion, Class A and D |
EN61000-3-3 | Line flicker EN60601-1-2, EN55024 |
EN61000-4-2 | ESD,±15 KV air and ±8 KV contact |
EN61000-4-3 | Radiated immunity, 10 V/m |
EN61000-4-4 | Fast transient/burst, ±2 KV |
EN61000-4-5 | Surge, ±1 KV diff. |
EN61000-4-6 | Conducted immunity, 10 Vrms |
EN61000-4-8 | Magnetic field immunity, 30 A/m |
EN61000-4-11 | Voltage dip immunity, 30% reduction for 500 ms, and >95% reduction for 10 ms |
PFD | TTL high for normal operation, low upon loss of input power, turn-on delay time 100-1000 ms, turn-off delay time 1 ms minimum |
Inhibit | Requires an external TTL high level signal to inhibit outputs for standard models |
Model
Model No. | I/P Vol. | O/P Vol. | O/P Current | O/P Watts (max.) |
FPM500-S120-z | 90-264VAC | 12V | 29.17A(Convection) / 37.50A(30 CFM) | 350W / 450W |
FPM500-S150-z | 90-264VAC | 15V | 23.34A(Convection) / 30.00A(30 CFM) | 350W / 450W |
FPM500-S180-z | 90-264VAC | 18V | 22.23A(Convection) / 27.78A(30 CFM) | 400W / 500W |
FPM500-S240-z | 90-264VAC | 24V | 16.67A(Convection) / 20.84A(30 CFM) | 400W / 500W |
FPM500-S280-z | 90-264VAC | 28V | 14.29A(Convection) / 17.86A(30 CFM) | 400W / 500W |
FPM500-S360-z | 90-264VAC | 36V | 11.12A(Convection) / 13.89A(30 CFM) | 400W / 500W |
FPM500-S480-z | 90-264VAC | 48V | 8.34A(Convection) / 10.42A(30 CFM) | 400W / 500W |
FPM500-S570-z | 90-264VAC | 57V | 7.02A(Convection) / 8.78A(30 CFM) | 400W / 500W |