80-100 Watt Single Output Medical Power Supplies : FPM101
FPM101 - 80-100 Watt Single Output Medical Power Supplies

Description
Medical Open Frame
The FPM101 series of AC-DC switching power supplies in a package of 2 x 4 x 1.29 inches are capable of delivering 100 watts of continuous power at 7.5 CFM forced air cooling or 80 watts at convection cooling. The units are constructed on a printed circuit board. They are suited for medical applications. Approval to both UL60601-1 safety standard improves design-in time and reduces end equipment compliance costs.Features
- BF class insulation
- Operation altitude up to 5000 meters
- 2 x 4 inch footprint with 1.29 inch low profile
- Less than 175 μA leakage current
- Wide input range 80-264 VAC
- Meet EN55011 /55032 and FCC Class B
- Short-circuit protection
- Compliant with RoHS requirements
- No load power consumption less than 0.15W
Specification
Input voltage | 80-264 VAC |
Power derating | Derate linearly from 100% at 90VAC To 90% at 85VAC and 80% at 80VAC |
Input frequency | 47-63 Hz |
Input current | 2.0 A (rms) for 115 VAC ; 1.2 A (rms) for 230 VAC |
Earth leakage current | 175 μA max. @ 264 VAC, 63 Hz |
Touch current | 100 μA max. @ 264 VAC, 63 Hz |
Ripple and noise | 1% peak to peak maximum |
Over voltage protection | set at 112-140% of its nominal output voltage, latching by recycle input to reset Short circuit protection:Automatic recovery Over temperature protection:Latching by recycle input to reset. |
Temperature coefficient | All outputs ±0.04% /℃ maximum |
Transient response | Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500 us after a 25% step load change |
Operating temperature | 20℃ to +70℃ |
Storage temperature | -40℃ to +85℃ |
Relative humidity | 5% to 95% non-condensing |
Temperature derating | Derate from 100% at +50℃ linearly to 50% at +70℃, applicable to convection and forced-air cooling conditions Switching frequency::65 KHz (typical) |
Efficiency | See rating chart. |
Hold-up time | 10 ms minimum at 100 W load and 115 VAC ; 10 ms minimum at 80 W load and 100 VAC |
Line regulation | ±0.5% maximum at full load |
Inrush current | 80 A @ 115 VAC or 160 A @ 230 VAC, at 25℃ cold start |
Withstand voltage | 4000 VAC from input to output, 1500 VAC from input to ground, 1500 VAC from output to ground |
MTBF | 150,000 hours at full load at 25℃ ambient, calculated per MIL-HDBK-217F EMC Performance |
EN55011 /EN55022 | Class B conducted, class B radiated |
EN61000-3-2 | Harmonic distortion, class A |
EN61000-3-3 | Line flicker |
EN61000-4-2 | ESD, ±15 KV air and ±8 KV contact |
EN61000-4-3 | Radiated immunity, 10 V/m |
EN61000-4-4 | Fast transient/burst, ±2 KV |
EN61000-4-5 | Surge, ±1 KV diff., ±2 KV com |
EN61000-4-6 | Conducted immunity, 10 Vrms |
EN61000-4-8 | Magnetic field immunity, 30 A/m |
EN61000-4-11 | Voltage dip immunity, 30% reduction for 500 ms, 100% reduction for 10 ms |
Model
Model No. | I/P Vol. | O/P Vol. | O/P Current | O/P Watts(max.) |
FPM101-S120 | 80-264VAC | 12VDC | 0~6.67A(convection)/8.34A(7.5CFM) | 80W / 100W |
FPM101-S150 | 80-264VAC | 15VDC | 0~5.34A(convection)/6.67A(7.5CFM) | 80W / 100W |
FPM101-S180 | 80-264VAC | 18VDC | 0~4.45A(convection)/5.56A(7.5CFM) | 80W / 100W |
FPM101-S240 | 80-264VAC | 24VDC | 0~3.34A(convection)/4.17A(7.5CFM) | 80W / 100W |
FPM101-S280 | 80-264VAC | 28VDC | 0~2.86A(convection)/3.58A(7.5CFM) | 80W / 100W |
FPM101-S320 | 80-264VAC | 32VDC | 0~2.50A(convection)/3.13A(7.5CFM) | 80W / 100W |
FPM101-S360 | 80-264VAC | 36VDC | 0~2.23A(convection)/2.78A(7.5CFM) | 80W / 100W |
FPM101-S480 | 80-264VAC | 48VDC | 0~1.67A(convection)/2.09A(7.5CFM) | 80W / 100W |