30-48 Watt Dual Output Medical Power Supplies : FPM042
FPM042 - Dual Output 30-48 Watt Medical Power Supplies
Description
Medical Open Frame
The FPM042 series of compact, open PCB constructed, AC-DC switching power supplies are capable of delivering 30-48 watts of continuous output power at convection cooling. They operate at 90-264 VAC input voltage without the need of voltage selection, and are suited for medical applications. Approval to UL60601-1 Safety Standard improves design-in time and reduces end equipment compliance costs. FPM042-zyyy x = S, D or T to denote multiple output ; S for single output, D for dual output, T for triple output, yyy to denote different output rating.Features
- BF Class insulation
- Medical approvals
- Compact size 2”×4”×1.18”
- Single, dual and triple outputs
- Wide-range input 90-264 VAC
- Low earth leakage current
- Level B emissions
- RoHS compliant
Specification
Input voltage | 90-264 VAC |
Input frequency | 47-63 Hz |
Input current | 0.9 A (rms) for 100 VAC ; 0.5 A (rms) for 240 VAC |
Earth Leakage current | 150 μA max. @ 264 VAC, 63 Hz |
Touch current | 100 μA max. @ 264 VAC, 63 Hz |
Output voltage/current | See rating chart. |
Maximum output power | See rating chart. |
Ripple and noise | 100 mV peak to peak on 3.3 V & 5.0 V models, 1% peak to peak on other models |
Overvoltage protection | Provided on output #1 only; set at 112–132% of its nominal output voltage |
Overcurrent protection | All outputs protected to short circuit conditions |
Temperature coefficient | All outputs ±0.04% /℃ maximum |
Transient response | Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500 us after a 25% step load change |
Operating temperature | -10℃ to +70℃ |
Storage temperature | -40℃ to +85℃ |
Relative humidity | 5% to 95% non-condensing |
Derating | Derate from 100% at +50℃ linearly to 50% at +70℃ |
Switching frequency | 62K ±5KHz |
Efficiency | 80-88% typical except FPM42-31-3A and FPM42-31-5A at 75% typical |
Hold-up time | 12 ms minimum at 110 VAC |
Line regulation | ±0.5% maximum at full load |
Inrush current | 25 A @ 115 VAC, or 50 A @ 230 VAC, at 25℃ cold start |
Withstand voltage | 4000 VAC from input to output (2 MOPP) ; 1500 VAC from input to ground (1 MOPP) ; 1500 VAC from output to ground |
MTBF | 400,000 hours at full load at 25℃ ambient, calculated per MIL-HDBK-217F EMC Performance |
EN55011/EN55022 | Class B conducted, class B radiated |
EN61000-3-2 | Harmonic distortion, class A and D |
EN61000-3-3 | Line flicker |
EN61000-4-2 | ESD, ±15 KV air and ±8 KV contact |
EN61000-4-3 | Radiated immunity, 10 V/m |
EN61000-4-4 | Fast transient/burst, ±2 KV |
EN61000-4-5 | Surge, ±1 KV diff., ±2 KV com |
EN61000-4-6 | Conducted immunity, 10 Vrms |
EN61000-4-8 | Magnetic field immunity, 30 A/m |
EN61000-4-11 | Voltage dip immunity, 30% reduction for 500 ms, 100% reduction for 10 ms |
Model
Model No. | I/P Vol. | O/P Vol. | O/P Current | O/P Watts(max.) |
FPM042-S050 | 90-264VAC | 5VDC | 0A~8.0A | 40 W |
FPM042-S120 | 90-264VAC | 12VDC | 0A~3.5A | 42 W |
FPM042-S150 | 90-264VAC | 15VDC | 0A~3.0A | 45 W |
FPM042-S240 | 90-264VAC | 24VDC | 0A~2.0A | 48 W |
FPM042-S480 | 90-264VAC | 48VDC | 0A~1.0A | 48 W |
FPM042-D023 | 90-264VAC | +5V / +12V | 0.5A~6.0A/0.1A~2.0A | 40 W |
FPM042-D025 | 90-264VAC | +5V / +24V | 0.5A~6.0A/0.1A~1.0A | 40 W |
FPM042-T031 | 90-264VAC | +5V / +12V / -12V | 0.5A~6.0A/0.1A~2.0A/0A~0.3A | 40 W |
FPM042-T031-3 | 90-264VAC | +3.3V / +5V / +12V | 0.8A~6.0A/0.1A~2.0A/0A~0.3A | 30 W |
FPM042-T031-5 | 90-264VAC | +5V / +3.3V / +12V | 0.5A~6.0A/0A~1.5A/0A~0.3A | 30 W |
FPM042-T032 | 90-264VAC | +5V / +15V / -15V | 0.5A~6.0A/0.1A~1.5A/0A~0.3A | 40 W |
FPM042-T039 | 90-264VAC | +5V / +24V / -12V | 0.5A~6.0A/0.1A~1.0A/0A~0.3A | 40 W |