FPM110 - 72~110 Watt Medical Triple Output Power Supplies, 72~110 Watt Medical Triple Output Power Supplies : FPM110
FranMar International Inc. continually develops the innovative FPM110 - 72~110 Watt Medical Triple Output Power Supplies to meet the various market demands and bring multiple benefits to customers always. Our FPM110 - 72~110 Watt Medical Triple Output Power Supplies are highly demanded by our clients across the country for optimum quality. In order to ensure longer service life and reliability, our products are manufactured by our vendors using the best quality raw material and latest technology. Whether you need a FPM110 - 72~110 Watt Medical Triple Output Power Supplies, or 72~110 WATT MEDICAL TRIPLE OUTPUT POWER SUPPLIES : FPM110, FPM110 - 72~110 WATT MEDICAL TRIPLE OUTPUT POWER SUPPLIES, FPM110 - 72~110 WATT Single Output MEDICAL POWER SUPPLIES, FranMar International Inc. manufactures cost-effective products.
72~110 Watt Medical Triple Output Power Supplies : FPM110
FPM110 - 72~110 Watt Medical Triple Output Power Supplies
Description :
Medical Open Frame
The FPM110 series of compact, open PCB constructed, AC-DC switching power supplies are specially designed for medical applications. They are capable of delivering 72-110 watts of continuous power at 25 CFM forced air cooling or 60-80 watts at convection cooling. They operate at 85-264 VAC input voltage without the need of a selector strap. All models meet the safety requirements of UL, CSA for medical equipment. FPM110-zyyy x = S, D, T or Q to denote multiple output ; S for single output, D for dual output, T for triple output, Q for Quad output, yyy to denote different output rating.Features :
- Low safety ground leakage current
- Meet EN55011 Class B
- Small size, light weigh
- 100% burn-in
- Wide input range 85-264VAC
- Input surge current protection
- Overvoltage protection
- Overcurrent protection
- Compliant with RoHS requirements
Specification :
Input voltage | 85-264 VAC |
Input frequency | 47-63 Hz |
Input current | 3.20 A (rms) for 115 VAC ; 1.80 A (rms) for 230 VAC |
Earth leakage current | 220 μA max. @ 264 VAC, 63 Hz |
Touch current | 100 μA max. @ 264 VAC, 63 Hz |
Output voltage/current | See rating chart. |
Maximum output power | See rating chart. |
Ripple and noise | 1% peak to peak maximum |
Overvoltage protection | Provided on output #1 only; set at 112-132% of its nominal output voltage |
Overcurrent protection | All outputs protected to short circuit conditions |
Temperature coefficient | All outputs ±0.04% /℃ maximum |
Transient response | Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500 us after a 25% step load change |
PFD | TTL logic high for normal operation and TTL logic low upon loss of input power. This signal appears at least 1 ms prior to V1 output dropping 5% below its nominal value. This signal also provides a minimum delay of 100 ms after V1 is within regulation |
Operating temperature | -10℃ to +70℃ |
Storage temperature | -40℃ to +85℃ |
Relative humidity | 5% to 95% non-condensing |
Derating | Derate from 100% at +50℃ linearly to 50% at +70℃ |
Cooling | 72-110 watts continuous output power at 25 CFM forced air cooling or 60-80 watts at convection cooling |
Switching frequency | 20-250 KHz, varied with load and line |
Efficiency | 70% minimum on single output model with Vo ≧12 V, 65% minimum on the others |
Hold-up time | 12 ms minimum at 110 VAC |
Line regulation | ±0.5% maximum at full load |
Inrush current | 15 A @ 115 VAC or 30 A @ 230 VAC, at 25℃ cold start |
Withstand voltage | 5600 VDC from input to output (2 MOPP) ; 2100 VDC from input to ground (1 MOPP) ; 700 VDC from output to ground(To verify AC strength, get correct test method to avoid power supply damage.)" |
MTBF | 400,000 hours at full load at 25℃ ambient, calculated per MIL-HDBK-217F EMC Performance (EN60601-1-2) |
EN55011 | Class B conducted, class B radiated |
EN61000-3-2 | Harmonic distortion, class A |
EN61000-3-3 | Line flicker |
EN61000-4-2 | ESD, ±15 KV air and ±8 KV contact |
EN61000-4-3 | Radiated immunity, 10 V/m |
EN61000-4-4 | Fast transient/burst, ±2 KV |
EN61000-4-5 | Surge, ±1 KV diff., ±2 KV com |
EN61000-4-6 | Conducted immunity, 10 Vrms |
EN61000-4-8 | Magnetic field immunity, 30 A/m |
EN61000-4-11 | Voltage dip immunity, 30% reduction for 500 ms, 100% reduction for 10 ms |
Model
Model No. | I/P Vol. | O/P Vol. | O/P Current | O/P Watts(max.) |
FPM110-S033 | 85-264VAC | 3.3V | 0~22A | 60W / 72W |
FPM110-S050 | 85-264VAC | 5.0V | 0~22A | 80W / 110W |
FPM110-S120 | 85-264VAC | 12V | 0~9.0A | 80W / 110W |
FPM110-S150 | 85-264VAC | 15V | 0~7.5A | 80W / 110W |
FPM110-S240 | 85-264VAC | 24V | 0~4.5A | 80W / 110W |
FPM110-S300 | 85-264VAC | 30V | 0~3.6A | 80W / 110W |
FPM110-D023 | 85-264VAC | +5.1V / +12V | 0~10A/0~5A/9.0A(Ipeak) | 80W / 110W |
FPM110-T031 | 85-264VAC | +5.1V / +12V / -12V | 0~10A/0~5A/9.0A(Ipeak)/0~1A | 80W / 110W |
FPM110-T032 | 85-264VAC | +5.1V / +15V / -15V | 0~10A/0~4A/7.5A(Ipeak)/0~1A | 80W / 110W |
FPM110-Q040 | 85-264VAC | +5.1V / +12V / -12V / -5V | 0~10A/0~5A/9.0A(Ipeak)/0~1A/0~1A | 80W / 110W |
FPM110-Q041 | 85-264VAC | +5.1V / +15V / -15V / +24V | 0~10A/0~4A/7.5A(Ipeak)/0~1A/0~1A | 80W / 110W |
FPM110-Q042 | 85-264VAC | +5.1V / +12V / -12V / +12V | 0~10A/0~5A/9.0A(Ipeak)/0~1A/0~1A | 80W / 110W |
FPM110-Q045 | 85-264VAC | +5.1V / +12V / -12V / +24V | 0~10A/0~5A/9.0A(Ipeak)/0~1A/0~1A | 80W / 110W |
FPM110-Q045-1 | 85-264VAC | +5.1V / +12V / -12V / +24V | 2~10A/0~5A/9.0A(Ipeak)/0~1A/1.5~3A | 80W / 110W |
FPM110-Q045-2 | 85-264VAC | +5.1V / +24V / -12V / +12V | 0~10A/0~3A/5.0A(Ipeak)/0~1A/0~1A | 80W / 110W |
FPM110-Q046 | 85-264VAC | +5.1V / +15V / -15V / -5V | 0~10A/0~4A/7.5A(Ipeak)/0~1A/0~1A | 80W / 110W |
FranMar International Inc. is the leading supplier of FPM110 - 72~110 Watt Medical Triple Output Power Supplies in Taiwan and the surrounding areas. We carry inventory all the time so you’re sure we’ll have your order ready whenever you need it. We can manufacture FPM110 - 72~110 Watt Medical Triple Output Power Supplies that meet specific technical requirements. FranMar International Inc. has tailored our customer service department, product designs, technical support as well as all other company departments to serve the ultimate goal of upholding the highest standards in customer satisfaction. We appreciate the opportunity to attract your attention and value your feedback.