200 WATT Single Output MEDICAL POWER SUPPLIES : FPM202

FPM202 - 200 WATT Single Output MEDICAL POWER SUPPLIES

Description

Medical Open Frame

The FPM202 series of AC-DC switching power supplies in a package of 3 x 5 x 1.5 inches are capable of delivering 200 watts of continuous power at 5.3 CFM forced air cooling or 150 watts at convection cooling. The units are constructed on a printed circuit board with a U-bracket for mechanical support and heat sinking. A cover-and-fan assembly can be added during manufacturing for 200 watt output. They are specially designed for medical applications. The units are UL/CSA 60601-1 certified. FPM202-Syyy-z yyy to denote different output rating ; z = U or E to denote different casing type ; U for U-bracket, E for Enclosed.
Features
  • BF Class insulation
  • Operation altitude up to 5000 meters
  • 3 x 5 inch footprint with 1.5 inch low profile
  • Less than 220 μA leakage current
  • Meet EN55011 /55022 and FCC Class B
  • Power Factor 0.98 typical
  • Short-circuit protection
  • Power Fail Detect (PFD) signal
  • Inhibit - TTL high to disable output
  • Compliant with RoHS requirements
  • High Efficiency 92% typical
Specification
Input voltage 90-264 VAC
Input frequency 47-63 Hz
Input current 2.5 A (rms) for 115 VAC ; 1.25 A (rms) for 230 VAC
Earth leakage current 220 μA max. @ 264 VAC, 63 Hz
Touch current 100 μA max. @ 264 VAC, 63 Hz
Output voltage/current See rating chart.
Total output power See rating chart.
Ripple and noise 1% peak to peak maximum
Remote sense Compensation for cable losses up to 0.5 V
Overvoltage protection Set at 112-140% of its nominal output voltage
Overcurrent protection Output protected to short circuit conditions
Temperature coefficient All outputs ±0.04% /℃ maximum
Transient response Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500 us after a 25% step load change
Fan power 12 V at 250 mA maximum
Operating temperature 0℃ to +70℃
Storage temperature -40℃ to +85℃
Relative humidity 5% to 95% non-condensing
Derating Derate from 100% at +50℃ linearly to 50% at +70℃, applicable to convection and forced-air cooling conditions
Switching frequency 100 KHz (typical)
Efficiency 87% minimum on all models
Hold-up time 10 ms minimum at 110 VAC
Line regulation ±0.5% maximum at full load
Inrush current 20 A @ 115 VAC or 40 A @ 230 VAC, at 25℃ cold start
Withstand voltage 4000 VAC from input to output (2MOPP) ; 1500 VAC from input to ground (1MOPP) ; 1500 VAC from output to ground
MTBF 300,000 hours at full load at 25℃ ambient, calculated per MIL-HDBK-217F
EMC Performance
EN55011 /EN55022 Class B conducted, class B radiated
EN61000-3-2 Harmonic distortion, class A and D
EN61000-3-3 Line flicker
EN61000-4-2 ESD, ±15 KV air and ±8 KV contact
EN61000-4-3 Radiated immunity, 10 V/m
EN61000-4-4 Fast transient/burst, ±2 KV
EN61000-4-5 Surge, ±1 KV diff., ±2 KV com
EN61000-4-6 Conducted immunity, 10 Vrms
EN61000-4-8 Magnetic field immunity, 30 A/m
EN61000-4-11 Voltage dip immunity, 30% reduction for 500 ms, 100% reduction for 10 ms

Model
Model No. I/P Vol. O/P Vol. O/P Current O/P Watts(max.)
FPM202-S120-z 90-264VAC 12V 0.1~12.50A(convection)/16.67A(5.3CFM) 150W/200W
FPM202-S150-z 90-264VAC 15V 0.1~10.00A(convection)/13.34A(5.3CFM) 150W/200W
FPM202-S180-z 90-264VAC 18V 0.1~8.34A(convection)/11.12A(5.3CFM) 150W/200W
FPM202-S240-z 90-264VAC 24V 0.1~6.25A(convection)/8.34A(5.3CFM) 150W/200W
FPM202-S280-z 90-264VAC 28V 0.1~5.36A(convection)/7.15A(5.3CFM) 150W/200W
FPM202-S320-z 90-264VAC 32V 0.1~4.69A(convection)/6.25A(5.3CFM) 150W/200W
FPM202-S360-z 90-264VAC 36V 0.1~4.17A(convection)/5.56A(5.3CFM) 150W/200W
FPM202-S480-z 90-264VAC 48V 0.1~3.13A(convection)/4.17A(5.3CFM) 150W/200W
Mechanical Drawing (Unit: mm)