400 WATT Single Output MEDICAL POWER SUPPLIES : FPM400
FPM400 - 400 WATT Single Output MEDICAL POWER SUPPLIES
Description
Medical Open Frame
The FPM400 series of AC-DC switching power supplies in a package of 4 x 7 x 1.58 inches are capable of delivering 400 watts of continuous power at 7 CFM forced air cooling or 300 watts at convection cooling. The units are constructed on a printed circuit board with a U-bracket for mechanical support and heat sinking. A cover and fan assembly can be added during manufacturing for 400 watt output without the change of any dimension. They are designed for medical applications.Features
- BF Class insulation
- Operation altitude up to 5000 meters
- 100-240 VAC input with active PFC
- Less than 300 μA leakage current
- Standby output 5VDC at 100mA
- EN55011 / 55022 Class B conducted emissions
- Inhibit - TTL low to disable output
- Standard PS Off and DC OK signals
- High Efficiency 92% typical
- Compliant with RoHS requirements
Specification
Input Voltage | 90-264 VAC |
Input frequency | 47-63 Hz |
Input Current | 4.2 A (rms) @115 VAC, 60 Hz ; 2.1 A (rms) @ 230 VAC, 50 Hz |
Earth leakage current | 300 μA max. @ 264 VAC, 63 Hz |
Touch current | 100 μA max. @ 264 VAC, 63 Hz |
Output voltage/current | See rating chart. |
Maximum output power | See rating chart. |
Ripple and noise | 1% peak to peak maximum |
Remote sense | Compensation for cable losses up to 0.5 V |
Overvoltage protection | Set at 115-140% of nominal output voltage |
Overcurrent protection | Protected to output short circuit conditions |
Thermal shutdown | Protected to overtemperature conditions |
Temperature coefficient | All outputs ±0.04% /℃ maximum |
Transient response | Maximum excursion of 4%, recovering to 1% of final value within 500 us after a 25% step load change |
Standby power | 5 V at 100 mA maximum |
Fan power | 12 V at 250 mA maximum |
Operating temperature | -10℃ to +70℃ |
Storage temperature | -40℃ to +85℃ |
Relative humidity | 5% to 95% non-condensing |
Derating | Derate from 100% at +50℃ linearly to 50% at +70℃, applicable to convection and forced-air cooling conditions |
Switching frequency | 85 KHz (typical) |
Efficiency | Typical 89% @ 115 VAC, 92% @ 230 VAC |
Hold-up time | 12 ms minimum at 110 VAC & 400 W |
Line regulation | ±0.5% maximum at full load |
Inrush current | 20 A @ 115 VAC, or 40 A @ 230 VAC, at 25℃ cold start |
Withstand voltage | 4000 VAC from input to output (2 MOPP) ; 1500 VAC from input to ground (1 MOPP) ; 1500 VAC from output to ground |
MTBF | 350,000 hours at full load at 25℃ ambient, calculated per MIL-HDBK-217F EMC Performance |
EN55011/ EN55022 | Class B conducted, Class A radiated |
EN61000-3-2 | Harmonic distortion, Class A and D |
EN61000-3-3 | Line flicker |
EN61000-4-2 | ESD,±15 KV air and ±8 KV contact |
EN61000-4-3 | Radiated immunity, 10 V/m |
EN61000-4-4 | Fast transient/burst, ±2 KV |
EN61000-4-5 | Surge, ±1 KV diff., ±2 KV com |
EN61000-4-6 | Conducted immunity, 10 Vrms |
EN61000-4-8 | Magnetic field immunity, 30 A/m |
EN61000-4-11 | Voltage dip immunity, 30% reduction for 500 ms, 100% reduction for 10 ms |
PFD | TTL logic high for normal operation and TTL logic low upon loss of input power. This signal appears at least 1 ms prior to master output dropping 5% below its nominal value. This signal also provides a minimum delay of 100 ms after master output is within regulation. |
Inhibit | TTL low to turn off output |
DC OK | TTL high when output voltage >95% |
PS OFF | TTL high to turn off output |
Model
Model No. | I/P Vol. | O/P Vol. | O/P Current | O/P Watts (max.) |
FPM400-S120-z | 90-264VAC | 12V | 0.1A~25.00A(Convection) / 33.34A(7 CFM) | 300W / 400W |
FPM400-S150-z | 90-264VAC | 15V | 0.1A~20.00A(Convection) / 26.67A(7 CFM) | 300W / 400W |
FPM400-S180-z | 90-264VAC | 18V | 0.1A~16.67A(Convection) / 22.23A(7 CFM) | 300W / 400W |
FPM400-S240-z | 90-264VAC | 24V | 0.1A~12.50A(Convection) / 16.67A(7 CFM) | 300W / 400W |
FPM400-S280-z | 90-264VAC | 28V | 0.1A~10.72A(Convection) / 14.29A(7 CFM) | 300W / 400W |
FPM400-S360-z | 90-264VAC | 36V | 0.1A~8.34A(Convection) / 11.12A(7 CFM) | 300W / 400W |
FPM400-S480-z | 90-264VAC | 48V | 0.1A~6.25A(Convection) / 8.34A(7 CFM) | 300W / 400W |