20-30W fixed AC plug Wallmount Type Waterproof Adapter : FWM030 Series
FWM030 - 20~30W Medical Power Supply with configuration versatility, can be used with different fixed AC plugs as needed.
Description
Medica Waterproofl Power
The FWM030 series of AC/DC wall mount switching power supplies are for 20-30 watts of continuous output power. They are enclosed in a 94V-0 rated polycarbonate case. The FWM030 models have fixed US and Euro AC plug with IP22. All models meet EN55011 class B emission limits, and are designed for medical applications.Features
- Operation altitude up to 5000 meters
- BF class insulation
- Meet IP22
- Overvoltage protection
- Over-temperature protection
- Short-circuit protection
- 100% burn-in at full rated load
- Compliant with DoE efficiency level VI requirements
- Compliant with RoHS requirements
Specification
Input Voltage | 80 ~ 264 VAC |
Input frequency | 47-63 Hz |
Input Current (rms) | 1.0 A (rms) for 115 VAC, 0.6 A (rms) for 230 VAC |
Touch current | 100μA max. @ 264 VAC, 63 Hz |
Output Voltage | 5 ~ 48 VDC |
Maximum output power | 20 ~ 30W |
Efficiency | 83.5 ~ 87.5% |
Ripple and noise | 100 ~ 480mV |
Overvoltage protection | 112% ~ 140% |
Overcurrent protection | Protected to short circuit conditions |
Temperature coefficient | ±0.04% /℃ maximum |
Transient response | Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500 us after a 25% step load change |
Operating temperature | -20℃ to +60℃ |
Storage temperature | -40℃ to +85℃ |
Relative humidity | 10% to 90% non-condensing |
Temperature derating | Derate from 100% at +40℃ linearly to 50% at +60℃ |
Hold-up time | 8 ms minimum at 115 VAC |
Turn on delay time | 3 s maximum at 115 VAC |
Efficiency | Compliant with DoE efficiency level VI requirements (see rating chart) |
Line regulation | ±1% maximum at full load |
Inrush current | 60 A @ 115 VAC or 120 A @ 230 VAC at 25℃ cold start |
Withstand Voltage | 4000 VAC from input to output |
MTBF | 300,000 hours at full load at 25℃ ambient, calculated per MIL-HDBK-217F EMC Performance (IEC 60601-1-2) |
EN55011 | Class B conducted, Class B radiated |
EN61000-3-2 | Harmonic distortion, Class A and D |
EN61000-3-3 | Line flicker |
EN61000-4-2 | ESD,±15 KV air and ±8 KV contact |
EN61000-4-3 | Radiated immunity, 10 V/m |
EN61000-4-4 | Fast transient/burst, ±2 KV |
EN61000-4-5 | Surge, ±1 KV diff. |
EN61000-4-6 | Conducted immunity, 10 Vrms |
EN61000-4-8 | Magnetic field immunity, 30 A/m |
EN61000-4-11 | Voltage dip immunity, 30% reduction for 500 ms, and >95% reduction for 10 ms |
Model
Model No. | I/P Vol. | O/P Vol. | O/P Current | O/P Watts (max.) | DoE 6 | CoC 5 |
FWM030-S050-x | 80-264 VAC | 5.00 VDC | 4.00A max. | 20W | VI | Tier 2 |
FWM030-S075-x | 80-264 VAC | 7.50 VDC | 3.00A max. | 22.5W | VI | Tier 2 |
FWM030-S090-x | 80-264 VAC | 9.00 VDC | 3.00A max. | 27W | VI | Tier 2 |
FWM030-S100-x | 80-264 VAC | 10.0 VDC | 2.70A max. | 27W | VI | Tier 2 |
FWM030-S120-x | 80-264 VAC | 12.0 VDC | 2.50A max. | 30W | VI | Tier 2 |
FWM030-S150-x | 80-264 VAC | 15.0 VDC | 2.00A max. | 30W | VI | Tier 2 |
FWM030-S180-x | 80-264 VAC | 18.0 VDC | 1.67A max. | 30W | VI | Tier 2 |
FWM030-S240-x | 80-264 VAC | 24.0 VDC | 1.25A max. | 30W | VI | Tier 2 |
FWM030-S300-x | 80-264 VAC | 30.0 VDC | 1.00A max. | 30W | VI | Tier 2 |
FWM030-S480-x | 80-264 VAC | 48.0 VDC | 0.63A max. | 30W | VI | Tier 2 |
Ordering Information:
- Note : Wall Mount AC plug Type,
x=U (US plug) ; x=E (Euro plug)