100-150 WATT Single Output MEDICAL POWER SUPPLIES : FPM150

FPM150 - 100-150 WATT Single Output MEDICAL POWER SUPPLIES

Description

Medical Open Frame

The FPM150 series of AC-DC switching power supplies in a package of 2 x 4 x 1.3 inches are capable of delivering 100-150 watts of continuous power at 7.5 CFM forced air cooling or 100 watts at convection cooling. The units are constructed on a printed circuit board. The units are certified also toUL/CSA 60601-1. They are specially designed for medical applications.
Features
  • BF Class insulation
  • Operation altitude up to 5000 meters
  • 2 x 4 inch footprint with 1.3 inch low profile
  • Less than 275 μA leakage current
  • Wide input range 80-264 VAC
  • Meet EN55011 /55022 Class B
  • Power Factor 0.98 typical
  • 100% burn-in at full load
  • Short-circuit protection
  • Over-temperature protection
  • Power Fail Detect (PFD) signal (optional)
  • Compliant with RoHS requirements
  • No load power consumption less than 0.5W without PFD or 1W with PFD
Specification
Input voltage 80-264 VAC
Input frequency 47-63 Hz
Input current 1.7 A (rms) for 115 VAC ; 0.85 A (rms) for 230 VAC
Earth leakage current 275 μA max. @ 264 VAC, 63 Hz
Touch current 100 μA max. @ 264 VAC, 63 Hz
Output voltage/current See rating chart.
Total output power See rating chart.
Ripple and noise See rating chart.
Remote sense Compensation for cable losses up to 0.5 V
Overvoltage protection set at 112-140% of its nominal output voltage
Overcurrent protection Output protected to short circuit conditions
Temperature coefficient All outputs ±0.04% /℃ maximum
Transient response Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500 us after a 25% step load change
Fan power 12 V at 0.5 A maximum (isolated)
Operating temperature 0℃ to +70℃
Storage temperature -40℃ to +85℃
Relative humidity 5% to 95% non-condensing
Derating Derate from 100% at +50℃ linearly to 50% at +70℃, applicable to convection and forced-air cooling conditions
Switching frequency 133 KHz (typical)
Efficiency See rating chart.
Hold-up time 10 ms minimum at 120 VAC
Line regulation ±0.5% maximum at full load
Inrush current 80 A @ 115 VAC or 160 A @ 230 VAC,at 25℃ cold start
Withstand voltage 4000 VAC from input to output (2 MOPP) ; 1500 VAC from input to ground (1 MOPP) ; 1500 VAC from output to ground
MTBF 250,000 hours at full load at 25℃ ambient, calculated per MIL-HDBK-217F
EMC Performance
EN55011/EN55022 Class B conducted, class B radiated
EN61000-3-2 Harmonic distortion, class A and D
EN61000-3-3 Line flicker
EN61000-4-2 ESD, ±15 KV air and ±8 KV contact
EN61000-4-3 Radiated immunity, 10 V/m
EN61000-4-4 Fast transient/burst, ±2 KV
EN61000-4-5 Surge, ±1 KV diff., ±2 KV com
EN61000-4-6 Conducted immunity, 10 Vrms
EN61000-4-8 Magnetic field immunity, 30 A/m
EN61000-4-11 Voltage dip immunity, 30% reduction for 500 ms, 100% reduction for 10 ms

Model
Model No. I/P Vol. O/P Vol. O/P Current O/P Watts(max.)
FPM150-S120-z 80-264VAC 12V 8.35A(Convection) / 12.50A(7.5 CFM) / 14.0A(Peak) 100W / 150W
FPM150-S150-z 80-264VAC 15V 6.70A(Convection) / 10.00A(7.5 CFM) / 11.0A(Peak) 100W / 150W
FPM150-S180-z 80-264VAC 18V 5.56A(Convection) / 8.34A(7.5 CFM) / 9.2A(Peak) 100W / 150W
FPM150-S240-z 80-264VAC 24V 4.20A(Convection) / 6.25A(7.5 CFM) / 7.0A(Peak) 100W / 150W
FPM150-S300-z 80-264VAC 30V 3.34A(Convection) / 5.00A(7.5 CFM) / 5.6A(Peak) 100W / 150W
FPM150-S360-z 80-264VAC 36V 2.78A(Convection) / 4.17A(7.5 CFM) / 4.6A(Peak) 100W / 150W
FPM150-S480-z 80-264VAC 48V 2.10A(Convection) / 3.13A(7.5 CFM) / 3.5A(Peak) 100W / 150W