200-300 WATT Single Output MEDICAL POWER SUPPLIES : FPM300
FPM300 - 200-300 WATT Single Output MEDICAL POWER SUPPLIES
Description
Medical Open Frame
The FPM300 series comprising single and multiple output models for 200-300 watts of continuous output power is specially designed for medical applications. They operate at 90-264 VAC input voltage without the need of a selector strap. All auxiliary outputs are with magnetic amplifier to keep regulation. The units are constructed on a printed circuit board with a U-bracket for mechanical support and heat sinking. A cover-and-fan assembly can be added during manufacturing. FPM300-zyyy-z x = S or Q to denote multiple output ; S for single output, Q for Quad output ; yyy to denote different output rating ; z = U or E to denote different casing type ; U for U-bracket, E for Enclosed.Features
- EN61000-3-2 class A and D compliant
- Power Factor 0.98 typical
- Overvoltage protection
- Short-circuit protection
- Power Fail Detect (PFD) signal
- 100% burn-in at full rated load
- Remote sense on output #1 and output #2
- Remote inhibit – TTL high to disable output
- Compliant with RoHS requirements
Specification
Input voltage | 90-264 VAC |
Input frequency | 47-63 Hz |
Input current | 4.7 A (rms) for 115 VAC ; 2.3 A (rms) for 230 VAC |
Earth leakage current | 300 μA max. @ 264 VAC, 63 Hz |
Touch current | 100 μA max. @ 264 VAC, 63 Hz |
Output voltage/current | See rating chart. |
Maximum output power | See rating chart. |
Ripple and noise | 2% peak to peak maximum on 3.3 V & 5.1 V and 1% peak to peak maximum on other voltage outputs |
Overvoltage protection | Provided on output #1 only; set at 115-140% of its nominal output voltage |
Overcurrent protection | All outputs protected to short circuit conditions |
Temperature coefficient | All outputs ±0.04% /℃ maximum |
Transient response | Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500 us after a 25% step load change |
Fan power | 12 V at 350 mA maximum for B version, 12 V at 100 mA maximum for C version |
PFD | TTL logic high for normal operation and TTL logic low upon loss of input power. This signal appears at least 1 ms prior to V1 output dropping 5% below its nominal value. This signal also provides a minimum delay of 100 ms after V1 output is within regulation. |
Inhibit | Requires an external TTL high level signal to inhibit outputs for standard models |
Operating temperature | 0℃ to +70℃ |
Storage temperature | -40℃ to +85℃ |
Relative humidity | 5% to 95% non-condensing |
Derating | Derate from 100% at +50℃, linearly to 50% at +70℃ |
Cooling | 200 /250 /300 watts continuous output power at 35 CFM forced air cooling or 100 /125 /150 watts at convention cooling |
Switching frequency | 70 KHz ±10 KHz |
Power factor | 0.98 typical |
Efficiency | 70% minimum on all models |
Hold-up time | 12 ms minimum at 110 VAC |
Line regulation | ±0.2% maximum at full load |
Inrush current | 30 A @ 115 VAC or 60 A @ 230 VAC, at 25℃ cold start |
Withstand voltage | 5600 VDC from input to output (2 MOPP) ; 2100 VDC from input to ground (1 MOPP) ; 700 VDC from output to ground(To verify AC strength, get correct test method to avoid power supply damage.) |
MTBF | 300,000 hours minimum at full load at 25℃ ambient, calculated per MIL-HDBK-217F EMC Performance (EN60601-1-2) |
EN55011 | Class B conducted, class B radiated |
EN61000-3-2 | Harmonic distortion, class A and D |
EN61000-3-3 | Line flicker |
EN61000-4-2 | ESD, ±15 KV air and ±8 KV contact |
EN61000-4-3 | Radiated immunity, 10 V/m |
EN61000-4-4 | Fast transient/burst, ±2 KV |
EN61000-4-5 | Surge, ±1 KV diff., ±2 KV com |
EN61000-4-6 | Conducted immunity, 10 Vrms |
EN61000-4-8 | Magnetic field immunity, 30 A/m |
EN61000-4-11 | Voltage dip immunity, 30% reduction for 500 ms, 100% reduction for 10 ms |
Model
Model No. | I/P Vol. | O/P Vol. | O/P Current | O/P Watts(max.) |
FPM300-S033-z | 90-264VAC | 3.3V | 3.0A~60.0A | 100W/200W |
FPM300-S051-z | 90-264VAC | 5.1V | 3.0A~60.0A | 150W/300W |
FPM300-S120-z | 90-264VAC | 12V | 1.2A~25.0A | 150W/300W |
FPM300-S150-z | 90-264VAC | 15V | 1.0A~20.0A | 150W/300W |
FPM300-S240-z | 90-264VAC | 24V | 0.6A~12.5A | 150W/300W |
FPM300-S300-z | 90-264VAC | 30V | 0.5A~10.0A | 150W/300W |
FPM300-S480-z | 90-264VAC | 48V | 0.5A~6.3A | 150W/300W |
FPM300-Q040-3-z | 90-264VAC | 3.3V / 5.1V / 12V / 12V | 3.0A~35.0A/2.0A~22A/0~4A/0~4A | 125W/250W |
FPM300-Q040-z | 90-264VAC | 5.1V / 12V / 12V / 5.1V | 2.0A~35.0A/1.0A~10A/0~4A/0~4A | 150W/300W |
FPM300-Q041-z | 90-264VAC | 5.1V / 15V / 15V / 24V | 2.0A~35.0A/0.8A~8A/0~4A/0~2.5A | 150W/300W |
FPM300-Q042-z | 90-264VAC | 5.1V / 12V / 12V / 12V | 2.0A~35.0A/1.0A~10A/0~4A/0~4A | 150W/300W |
FPM300-Q045-z | 90-264VAC | 5.1V / 12V / 12V / 24V | 2.0A~35.0A/1.0A~10A/0~4A/0~2.5A | 150W/300W |
FPM300-Q046-z | 90-264VAC | 5.1V / 12V / 12V / 15V | 2.0A~35.0A/1.0A~10A/0~4A/0~4A | 150W/300W |
FPM300-Q047-z | 90-264VAC | 5.1V / 24V / 12V / 12V | 2.0A~35.0A/0.5A~5A/0~4A/0~4A | 150W/300W |
FPM300-Q048-z | 90-264VAC | 5.1V / 24V / 5.1V / 15V | 2.0A~35.0A/0.5A~5A/0~4A/0~4A | 150W/300W |
FPM300-Q049-z | 90-264VAC | 5.1V / 12V / 5.1V / 24V | 2.0A~35.0A/1.0A~10A/0~4A/0~2.5A | 150W/300W |
FPM300-Q410-z | 90-264VAC | 24V / 12V / 5.1V / 12V | 0.5A~6.3A/1.0A~10A/0~4A/0~4A | 150W/300W |
FPM300-Q411-z | 90-264VAC | 24V / 12V / 5.1V / 24V | 0.5A~6.3A/1.0A~10A/0~4A/0~2.5A | 150W/300W |
FPM300-Q412-z | 90-264VAC | 24V / 12V / 12V / 12V | 0.5A~6.3A/1.0A~10A/0~4A/0~4A | 150W/300W |
FPM300-Q413-z | 90-264VAC | 24V / 24V / 5.1V / 15V | 0.5A~6.3A/0.5A~5A/0~4A/0~4A | 150W/300W |
FPM300-Q414-z | 90-264VAC | 24V / 24V / 12V / 12V | 0.5A~6.3A/0.5A~5A/0~4A/0~4A | 150W/300W |