37.5-64 Watt Medical Triple Output Power Supplies : FPM060
FPM060 - 37.5-64 Watt Medical Triple Output Power Supplies

Description
Medical Open Frame
The FPM060 series of compact, open PCB constructed, AC-DC switching power supplies are capable of delivering 37.5-64 watts of continuous output power at convection cooling. They operate at 90-264 VAC input voltage without the need of voltage selection, and are suited for medical applications. Approval to UL60601-1 safety standards improves design-in time and reduces end equipment compliance costs. FPM060-zyyy x = S, D or T to denote multiple output ; S for single output, D for dual output, T for triple output, yyy to denote different output rating.Features
- BF Class insulation
- Medical approvals
- Compact size 2”×4”×1.18”
- Single, dual and triple outputs
- Wide-range input 90-264 VAC
- Low earth leakage current
- Level B emissions
- RoHS compliant
Specification
Input voltage | 90-264 VAC |
Input frequency | 47-63 Hz |
Input current | 1.3 A (rms) for 100 VAC ; 0.7 A (rms) for 240 VAC |
Earth Leakage current | 150 μA max. @ 264 VAC, 63 Hz |
Touch current | 100 μA max. @ 264 VAC, 63 Hz |
Output voltage/current | See rating chart. |
Maximum output power | See rating chart. |
Ripple and noise | 100 mV peak to peak on 3.3 V & 5.0 V models, 1% peak to peak on other models |
Overvoltage protection | Provided on output #1 only; set at 112–132% of its nominal output voltage |
Overcurrent protection | All outputs protected to short circuit conditions |
Temperature coefficient | All outputs ±0.04% /℃ maximum |
Transient response | Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500 us after a 25% step load change |
Operating temperature | -10℃ to +70℃ |
Storage temperature | -40℃ to +85℃ |
Relative humidity | 5% to 95% non-condensing |
Derating | Derate from 100% at +50℃ linearly to 50% at +70℃ |
Switching frequency | 62K ±5KHz |
Efficiency | 80-88% typical except FPM60-31-3A and FPM60-31-5A at 75% typical |
Hold-up time | 12 ms minimum at 110 VAC |
Line regulation | ±0.5% maximum at full load |
Inrush current | 30 A @ 115 VAC, or 60 A @ 230 VAC, at 25℃ cold start |
Withstand voltage | 4000 VAC from input to output (2 MOPP) ; 1500 VAC from input to ground (1 MOPP) ; 1500 VAC from output to ground |
MTBF | 400,000 hours at full load at 25℃ ambient, calculated per MIL-HDBK-217F EMC Performance |
EN55011 /EN55022 | Class B conducted, class B radiated |
EN61000-3-2 | Harmonic distortion, class A and D |
EN61000-3-3 | Line flicker |
EN61000-4-2 | ESD, ±15 KV air and ±8 KV contact |
EN61000-4-3 | Radiated immunity, 10 V/m |
EN61000-4-4 | Fast transient/burst, ±2 KV |
EN61000-4-5 | Surge, ±1 KV diff., ±2 KV com |
EN61000-4-6 | Conducted immunity, 10 Vrms |
EN61000-4-8 | Magnetic field immunity, 30 A/m |
EN61000-4-11 | Voltage dip immunity, 30% reduction for 500 ms, 100% reduction for 10 ms |
Model
Model No. | I/P Vol. | O/P Vol. | O/P Current | O/P Watts(max.) |
FPM060-S050 | 90-264VAC | 5VDC | 0A~11.0A | 55W |
FPM060-S120 | 90-264VAC | 12VDC | 0A~5.0A | 60W |
FPM060-S150 | 90-264VAC | 15VDC | 0A~4.3A | 64W |
FPM060-S240 | 90-264VAC | 24VDC | 0A~2.7A | 64W |
FPM060-S480 | 90-264VAC | 48VDC | 0A~1.35A | 64W |
FPM060-D023 | 90-264VAC | +5V / +12V | 0.5~6.0A(convection)/8A(5CFM)0.1~3.0A | 55W |
FPM060-D025 | 90-264VAC | +5V / +24V | 0.5~6.0A(convection)/8A(5CFM)0.1~1.5A | 55W |
FPM060-T031 | 90-264VAC | +5V / +12V / -12V | 0.5~6.0A(convection)/8A(5CFM)/0.1~3.0A/0~0.5A | 55W |
FPM060-T031-5 | 90-264VAC | +5V / +3.3V / +12V | 0.5~6.0A(convection)/8A(5CFM)/0~1.5A/0~0.5A | 37.5W |
FPM060-T032 | 90-264VAC | +5V / +15V / -15V | 0.5~6.0A(convection)/8A(5CFM)/0.1~2.4A/0~0.5A | 55W |
FPM060-T039 | 90-264VAC | +5V / +24V / -12V | 0.5~6.0A(convection)/8A(5CFM)/0.1~1.5A/0~0.5A | 55W |