96~120W Medical Waterproof AC Adapter : FWM120 Series
FWM120 - 96~120W AC/DC Medical Waterproof power adapter with 4 kind AC inlet available for world-wide use designed for medical applications.
Description
Waterproof Medical DeskTop Adapter
The FWM120 series of AC/DC switching power supplies are for 96-120 watts of continuous output power. They are enclosed in a 94V-0 rated polycarbonate case with IP54 level which have IEC320/C14, C6, C8 or C18 inlet to mate with interchangeable cord for world-wide use. All models meet EN55011 class B emission limits, and are designed for medical applications.Features
- No load power consumption less than 0.15W
- Meet Energy Start EPS2.0 /ErP EC No 278/2009 (Lot 7)
- Meet EU CoC EPS V5 Tier 2
- With PFC circuit
- Low profile at 28mm
- Operating altitude up to 5000 meters
- Wide input range 80 to 264 VAC
- Class II model meet IEC 60601-1-11 and IP54
- Optional output connectors
- 100% burn-in
- Overvoltage protection
- Overcurrent protection
- Compliant with RoHS requirements
Specification
Input Voltage | 80 ~ 264 VAC |
Power derating | Derate linearly from 100% at 90 VAC to 90% at 85Vac and 80% at 80 VAC |
Input frequency | 47-63 Hz |
Input Current (rms) | 1.60 A (rms) for 115 VAC, 0.80 A (rms) for 230 VAC |
Earth leakage current | 180 μA max. @ 264 VAC, 63 Hz |
Touch current | 100μA max. @ 264 VAC, 63 Hz |
Output Voltage | 12 ~ 48 VDC |
Maximum output power | 96 ~ 120W |
Ripple and noise | 1% peak to peak maximum |
Over voltage protection | 112% ~ 140% |
Short circuit protection | Automatic recovery |
Over temperature protection | Latching by recycle input to reset |
Temperature coefficient | ±0.04% /℃ maximum |
Transient response | Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500 us after a 25% step load change |
Operating temperature | -20℃ to +60℃ |
Atmospheric pressure | 540 hPa to 1060 hPa |
Storage temperature | -40℃ to +85℃ |
Relative humidity | 5% to 95% non-condensing |
Temperature derating | Derate from 100% at +40℃ linearly to 50% at +60℃ |
Ingress protection | IP54 |
Switching frequency | 50-130 KHz |
Power factor | 0.98 Typical at 115 VAC |
Efficiency | 88% min. at full load |
Hold-up time | 20 ms minimum at 100 VAC |
Line regulation | ±0.5% maximum at full load |
Inrush current | 90 A @ 115 VAC or 180 A @ 230 VAC, at 25℃ cold start |
Withstand voltage | 4000 VAC from input to output (2 MOPP) 1500 VAC from input to ground (1 MOPP) 500 VAC from output to ground (For class II models, 4000VAC from input to output) |
MTBF | 100,000 hours at full load at 25℃ambient , calculated per MIL-HDBK-217F EMC Performance (IEC60601-1-2) |
EN55011 | Class B conducted, class B radiated |
EN61000-3-2 | Harmonic distortion, class A and D |
EN61000-3-3 | Line flicker |
EN61000-4-2 | ESD, ±15 KV air and ±8 KV contact |
EN61000-4-3 | Radiated immunity, 10 V/m |
EN61000-4-4 | Fast transient/burst, ±2 KV |
EN61000-4-5 | Surge, ±1 KV diff., ±2 KV com. |
EN61000-4-6 | Conducted immunity, 10 Vrms |
EN61000-4-8 | Magnetic field immunity, 30 A/m |
EN61000-4-11 | Voltage dip immunity, 30% reduction for 500 ms (criteria A @ 230VAC, criteria B @ 100VAC), 60% reduction for 100 ms (criteria A @ 230VAC, criteria B @ 100VAC) and >95% reduction for 20 ms |
Model
Model No. | I/P Vol. | O/P Vol. | O/P Current | O/P Watts (max.) | DoE 6 | CoC 5 |
FWM120-S120-x | 80~264 VAC | 12.0VDC | 8.00A | 96W | VI | Tier 2 |
FWM120-S150-x | 80~264 VAC | 15.0VDC | 7.00A | 105W | VI | Tier 2 |
FWM120-S180-x | 80~264 VAC | 18.0VDC | 6.67A | 120W | VI | Tier 2 |
FWM120-S190-x | 80~264 VAC | 19.0VDC | 6.32A | 120W | VI | Tier 2 |
FWM120-S200-x | 80~264 VAC | 20.0VDC | 6.00A | 120W | VI | Tier 2 |
FWM120-S240-x | 80~264 VAC | 24.0VDC | 5.00A | 120W | VI | Tier 2 |
FWM120-S300-x | 80~264 VAC | 30.0VDC | 4.00A | 120W | VI | Tier 2 |
FWM120-S360-x | 80~264 VAC | 36.0VDC | 3.34A | 120W | VI | Tier 2 |
FWM120-S480-x | 80~264 VAC | 48.0VDC | 2.50A | 120W | VI | Tier 2 |
Ordering Information:
- Note : x for AC Inlet Type. It can be 4, 5, 6, or 8
4 = IEC C14 ; 5 = IEC C18 ; 6 = IEC C6 ; 8 = IEC C8