FranMar International Inc. continually develops the innovative FPM1100 - 1100 Watt Single Output AC DC Enclosed Medical Power Supplies to meet the various market demands and bring multiple benefits to customers always. Our FPM1100 - 1100 Watt Single Output AC DC Enclosed Medical Power Supplies are highly demanded by our clients across the country for optimum quality. In order to ensure longer service life and reliability, our products are manufactured by our vendors using the best quality raw material and latest technology. Whether you need a FPM1100 - 1100 Watt Single Output AC DC Enclosed Medical Power Supplies, or 1100 WATT Single Output MEDICAL POWER SUPPLIES : FPM1100, FPM1100 - 1100 WATT Single Output AC DC ENCLOSED MEDICAL POWER SUPPLIES, FPM1100 - 1100 WATT Single Output AC DC ENCLOSED MEDICAL POWER SUPPLIES, FranMar International Inc. manufactures cost-effective products.
1100 Watt Single Output Medical Power Supplies : FPM1100
FPM1100 - 1100 Watt Single Output AC DC Enclosed Medical Power Supplies
Description :
Medical Open Frame
The FPM1100 series of AC-DC switching medical power supplies in a package of 5.91 x 9.25 x 2.4 inches are capable of delivering 1100 watts of continuous power. The units are constructed on a printed circuit board with an enclosed format for mechanical support and heat sinking. They are designed for medical applications including those needing BF rated insulation and/or an operation altitude up to 5000 meters.Features :
- BF Class insulation
- Operation altitude up to 5000 meters
- Compact size 5.91” x 9.25” x 2.4”
- Less than 300 μA leakage current
- EN55011 Class B conducted emissions
- Inhibit - TTL low to disable output
- Standard PS Off and DC OK signals
- High Efficiency 89% typical
- Compliant with RoHS requirements
- Standby output 5 VDC at 200 mA
- Variable speed internal fan
- Overvoltage protection
- Overcurrent protection
- Thermal protection
Specification :
Input Voltage | 90-264 VAC |
Input frequency | 47-63 Hz |
Input Current (rms) | 16 A (rms) @100 VAC, 60 Hz; 8 A (rms) @ 240 VAC, 50 Hz |
Earth leakage current | 300 μA max. @ 264 VAC, 63 Hz |
Touch current | 100 μA max. @ 264 VAC, 63 Hz |
Ripple and noise | 1% peak to peak maximum Remote sense Compensation for cable losses up to 0.5 V |
Overvoltage protection | Set at 112-140% of nominal output voltage |
Overcurrent protection | Set at 120-140% of maximum output current Thermal shutdown Protected to overtemperature conditions |
Temperature coefficient | All outputs ±0.04% /℃ maximum |
Transient response | Maximum excursion of 4%, recovering to 1% of final value within 500 us after a 25% step load change Standby power 5 V at 200 mA maximum Fan power 12 V at 1.0 A maximum |
Operating temperature | -10℃ to +70℃ |
Storage temperature | -40℃ to +85℃ |
Relative humidity | 5% to 95% non-condensing |
Derating | Derate from 100% at +50℃ linearly to 50% at +70℃, applicable to convection and forced-air cooling conditions Switching frequency: 40 KHz to 200 KHz |
Efficiency | 88% min. |
Hold-up time | 10 ms minimum at 110 VAC |
Line regulation | ±0.5% maximum at full load |
Inrush current | 50 A @ 115 VAC, or 100 A @ 230 VAC, at 25℃ cold start |
Withstand voltage | 4000 VAC from input to output (2MOPP); 1500 VAC from input to ground (1MOPP); 1500 VAC from output to ground |
MTBF | 300,000 hours at full load at 25℃ ambient, calculated per MIL-HDBK-217F EMC Performance (IEC60601-1-2) |
EN55011 | Class B conducted, Class B radiated |
EN61000-3-2 | Harmonic distortion, Class A and D |
EN61000-3-3 | Line flicker |
EN61000-4-2 | ESD,±15 KV air and ±8 KV contact |
EN61000-4-3 | Radiated immunity, 10 V/m |
EN61000-4-4 | Fast transient/burst, ±2 KV |
EN61000-4-5 | Surge, ±1 KV diff. |
EN61000-4-6 | Conducted immunity, 10 Vrms |
EN61000-4-8 | Magnetic field immunity, 30 A/m |
EN61000-4-11 | Voltage dip immunity, 30% reduction for 500 ms, and >95% reduction for 10 ms |
PFD | TTL high for normal operation,low upon loss of input power, turn-on delay time 100-2500 ms, turn-off delay time 1 ms minimum |
Inhibit | TTL low to turn off output |
DC OK | TTL high when output voltage >95% |
PS OFF | TTL high to turn off output |
Model
Model No. | I/P Vol. | O/P Vol. | O/P Current | O/P Watts (max.) |
FPM1100-S240-E | 90-264VAC | 24 V | 45.84A / 52.1 A(Peak) | 1100W/1250W |
FPM1100-S280-E | 90-264VAC | 28 V | 39.29A / 44.65 A(Peak) | 1100W/1250W |
FPM1100-S320-E | 90-264VAC | 32 V | 34.38A / 39.07 A(Peak) | 1100W/1250W |
FPM1100-S340-E | 90-264VAC | 34 V | 32.35A / 36.77 A(Peak) | 1100W/1250W |
FPM1100-S360-E | 90-264VAC | 36 V | 30.56A / 34.73 A(Peak) | 1100W/1250W |
FPM1100-S420-E | 90-264VAC | 42V | 26.2A / 29.77A(Peak) | 1100W/1250W |
FPM1100-S480-E | 90-264VAC | 48 V | 22.92A / 26.1 A(Peak) | 1100W/1250W |
FranMar International Inc. is the leading supplier of FPM1100 - 1100 Watt Single Output AC DC Enclosed Medical Power Supplies in Taiwan and the surrounding areas. We carry inventory all the time so you’re sure we’ll have your order ready whenever you need it. We can manufacture FPM1100 - 1100 Watt Single Output AC DC Enclosed Medical Power Supplies that meet specific technical requirements. FranMar International Inc. has tailored our customer service department, product designs, technical support as well as all other company departments to serve the ultimate goal of upholding the highest standards in customer satisfaction. We appreciate the opportunity to attract your attention and value your feedback.